Micromega Heroshaper Classic Inget Files

Inhouse product


Price
₹1,900.00 ₹3,000.00 /kg 37% OFF
Length
Size
Quantity
(140 available)
Total Price
21mm-(20-4%,20-6%,25-4%,25-6%,30-4%,30-6%)

Micromega Heroshaper Classic Inget Files 21mm - (20 - 4%,20 - 6%,25 - 4%,25 - 6%,30 - 4%,30 - 6%)

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-20-4%

Micromega Heroshaper Classic Inget Files 21mm - 20 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-20-6%

Micromega Heroshaper Classic Inget Files 21mm - 20 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-25-4%

Micromega Heroshaper Classic Inget Files 21mm - 25 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-25-6%

Micromega Heroshaper Classic Inget Files 21mm - 25 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-30-4%

Micromega Heroshaper Classic Inget Files 21mm - 30 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
21mm-30-6%

Micromega Heroshaper Classic Inget Files 21mm - 30 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-(20-4%,20-6%,25-4%,25-6%,30-4%,30-6%)

Micromega Heroshaper Classic Inget Files 25mm - (20 - 4%,20 - 6%,25 - 4%,25 - 6%,30 - 4%,30 - 6%)

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-20-4%

Micromega Heroshaper Classic Inget Files 25mm - 20 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-20-6%

Micromega Heroshaper Classic Inget Files 25mm - 20 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-25-4%

Micromega Heroshaper Classic Inget Files 25mm - 25 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-25-6%

Micromega Heroshaper Classic Inget Files 25mm - 25 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-30-4%

Micromega Heroshaper Classic Inget Files 25mm - 30 - 4%

Price: ₹1,900.00 ₹3,000.00 37% OFF
25mm-30-6%

Micromega Heroshaper Classic Inget Files 25mm - 30 - 6%

Price: ₹1,900.00 ₹3,000.00 37% OFF

Total Price : ₹0.00

Items : 0

Share Now
Top Selling Products

Hero Shaper files, with their pronounced tapers, are designed for use with the crown-down technique to progressively remove constraints and flare the canal. The coronal third is prepared using .06 taper Hero Shaper files and the apical third is prepared using .04 taper Hero Shaper files. The varying helical pitch and length of the cutting portion of the files provide them with an excellent combination of efficiency and flexibility. The main advantage of this file method is that it requires a reduced number of instruments to prepare a canal 2, 3 or 4 depending on the difficulty. This approach simplifies the process, reduces intervention time and minimizes the risks associated with the use of too many instruments, while maintaining precise and efficient canal preparation.

InGeT (Integrated Gear Technology) is a specific coupling concept between a file and a contra-angle, with the incorporation of the driving gear into the rotary file itself. 

Benefits of the InGeT system Very compact head: 6.5 mm diameter and 7.5 mm high.

  • Increased visibility of the operative field.
  • Better visibility of the instrument when it is inserted in the canal and a better vision of the instrument while working. This is particularly useful when treatment is made under magnification (magnifying loupe or microscope).
  • Considerable reduction of the total length of the instrument + contra-angle head in the mouth.
  • Exceptional working comfort.

Innovative Pitch- Avoids screwing effects: the helical angle of the cutting edges varies from the tip to the shank. The new pitch also varies according to the taper: the more tapered an instrument is, the longer is its pitch; this can be described as an ADAPTED PITCH. Increased instrument performance: Efficiency / Flexibility / Strength

File Design-The blade shows a triple helix cutting edge. The tip is inactive in order to follow the canal anatomy. Positive rake angle for excellent debris evacuation. Big inner core for excellent resistance to breakage. 3 edges for better force balance. Increased safety with no screwing effect and excellent debris evacuation.

Anti-Breakage Control: The instrument “unwinds” to prevent the risk of fracture.

    • Maximum Torque- 1.5 Ncm
    • Rotation Speed- 400 rpm
    • Taper- 4% & 6%
    • 1 x Pack of 6 files

    According to canal type

    Rotation speed

    • The rotation speed must be 300 to 600 rpm.

    First penetration

    • The first penetration is performed with a conventional hand instrument (usually a n°10 K file-MMC n°10/21 mm). This is used to confirm the working length (WL), canal anatomy and diagnosis from the pre-operative X-ray.
    • Canal mineralisation and curvature will determine the treatment difficulty and thus the most suitable instrumentation sequence for the clinical case.

    Penetration levels

    • 06 taper files are brought to 2/3 of the working length. 04 taper files are brought to Working Length. An accurate determination of the working length between the passage of the .06 and .04 file is strongly recommended.

    Operating procedure

    • Hero Shaper is positioned at the canal entrance. The motor is started and maintained at constant speed whilst the file is working and until it is completely withdrawn from the canal.
    • The file is inserted to the desired working length with a short in and out movement using light pressure.
    • Hero Shaper also allows circumferential filing to be achieved which permits the canal anatomy to be taken into account and in particular to shape lateral zones.

    Irrigation

    • It is important to thoroughly irrigate the canal between the passage of successive instruments.

    Yellow Sequence (Complex Cases)

    1. Bring the HERO Shaper® n°20 - .06 taper to the 2/3 of Working Length.
    2. Bring the HERO Shaper® n°20 - .04 taper to Woking Length.
    3. Bring the HERO Shaper® n°25 - .04 taper to Working Length.
    4. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

    Red Sequence (Intermediate Cases)

    1. Bring the HERO Shaper® n°25 - .06 taper to the 2/3 of Working Length.
    2. Bring the HERO Shaper® n°25 - .04 taper to Working Length.
    3. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

    Blue Sequence (Simple Cases)

    1. Bring the HERO Shaper® n°25 - .06 taper to the 2/3 of Working Length.
    2. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

    • Country of Origin – Made in France
    • Manufacturer- Micromega

    Hero Shaper files, with their pronounced tapers, are designed for use with the crown-down technique to progressively remove constraints and flare the canal. The coronal third is prepared using .06 taper Hero Shaper files and the apical third is prepared using .04 taper Hero Shaper files. The varying helical pitch and length of the cutting portion of the files provide them with an excellent combination of efficiency and flexibility. The main advantage of this file method is that it requires a reduced number of instruments to prepare a canal 2, 3 or 4 depending on the difficulty. This approach simplifies the process, reduces intervention time and minimizes the risks associated with the use of too many instruments, while maintaining precise and efficient canal preparation.

    InGeT (Integrated Gear Technology) is a specific coupling concept between a file and a contra-angle, with the incorporation of the driving gear into the rotary file itself. 

    Benefits of the InGeT system Very compact head: 6.5 mm diameter and 7.5 mm high.

    • Increased visibility of the operative field.
    • Better visibility of the instrument when it is inserted in the canal and a better vision of the instrument while working. This is particularly useful when treatment is made under magnification (magnifying loupe or microscope).
    • Considerable reduction of the total length of the instrument + contra-angle head in the mouth.
    • Exceptional working comfort.

    Innovative Pitch- Avoids screwing effects: the helical angle of the cutting edges varies from the tip to the shank. The new pitch also varies according to the taper: the more tapered an instrument is, the longer is its pitch; this can be described as an ADAPTED PITCH. Increased instrument performance: Efficiency / Flexibility / Strength

    File Design-The blade shows a triple helix cutting edge. The tip is inactive in order to follow the canal anatomy. Positive rake angle for excellent debris evacuation. Big inner core for excellent resistance to breakage. 3 edges for better force balance. Increased safety with no screwing effect and excellent debris evacuation.

    Anti-Breakage Control: The instrument “unwinds” to prevent the risk of fracture.

      • Maximum Torque- 1.5 Ncm
      • Rotation Speed- 400 rpm
      • Taper- 4% & 6%

      • 1 x Pack of 6 files

      According to canal type

      Rotation speed

      • The rotation speed must be 300 to 600 rpm.

      First penetration

      • The first penetration is performed with a conventional hand instrument (usually a n°10 K file-MMC n°10/21 mm). This is used to confirm the working length (WL), canal anatomy and diagnosis from the pre-operative X-ray.
      • Canal mineralisation and curvature will determine the treatment difficulty and thus the most suitable instrumentation sequence for the clinical case.

      Penetration levels

      • 06 taper files are brought to 2/3 of the working length. 04 taper files are brought to Working Length. An accurate determination of the working length between the passage of the .06 and .04 file is strongly recommended.

      Operating procedure

      • Hero Shaper is positioned at the canal entrance. The motor is started and maintained at constant speed whilst the file is working and until it is completely withdrawn from the canal.
      • The file is inserted to the desired working length with a short in and out movement using light pressure.
      • Hero Shaper also allows circumferential filing to be achieved which permits the canal anatomy to be taken into account and in particular to shape lateral zones.

      Irrigation

      • It is important to thoroughly irrigate the canal between the passage of successive instruments.

      Yellow Sequence (Complex Cases)

      1. Bring the HERO Shaper® n°20 - .06 taper to the 2/3 of Working Length.
      2. Bring the HERO Shaper® n°20 - .04 taper to Woking Length.
      3. Bring the HERO Shaper® n°25 - .04 taper to Working Length.
      4. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

      Red Sequence (Intermediate Cases)

      1. Bring the HERO Shaper® n°25 - .06 taper to the 2/3 of Working Length.
      2. Bring the HERO Shaper® n°25 - .04 taper to Working Length.
      3. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

      Blue Sequence (Simple Cases)

      1. Bring the HERO Shaper® n°25 - .06 taper to the 2/3 of Working Length.
      2. Bring the HERO Shaper® n°30 - .04 taper to Working Length.

      • Country of Origin – Made in France
      • Manufacturer- Micromega